Settings - Scalar Measurement Results

Settings and scalar measurement results are indicated in the two parameter lines above the test diagram.

1st Parameter Line

The first parameter line contains the following settings:

RF Max. Level Maximum value of input level as set in the Connection Control menu (see section Table Oriented Version - Analyzer Settings).
RF Attenuation Setting for the attenuation of the input level (Normal, Low Noise, Low Distortion) as set in the Connection Control menu (see section Table Oriented Version - Analyzer Settings).
Freq. Center frequency of the RF signal analyzed
Trigger Trigger mode (Free Run, RF Power, IF Power or External)
 

2nd Parameter Line

The second parameter line contains the following marker values:

Level and time of reference marker
   
Level and time of delta marker 1 (setting absolute) or difference from reference marker (setting relative)
   
Level and time of delta marker 2 (setting absolute) or difference from reference marker (setting relative)
 

Remote control
Settings are read out using the query corresponding to the setting command (setting command with appended question mark).
To obtain the measurement value at a single point on the trace the whole trace must be read, see below.