Home Previous Next


Bandwidth (DC Coup.)

CONFigure:AFANalyzer[:PRIMary]:FILTer:BPASs:DCCoupling <Band pass>

CONFigure:AFANalyzer:SECondary:FILTer:BPASs:DCCoupling <Band pass>

<Band pass>

Description of parameters

Def. value

Def. unit


BP02 |
BP03 |
BP05 |
BP06 |
BP07 |
BP09 |
BP10 |
BP11 |
BP13 |
BP14 |
BP15 |
BP17 |
BP18

CMU band pass filter with a bandwidth of
6 Hz to 250 Hz
50 Hz to 250 Hz
6 Hz to 3000 Hz
50 Hz to 3000 Hz
300 Hz to 3000 Hz
6 Hz to 4000 Hz
50 Hz to 4000 Hz
300 Hz to 4000 Hz
6 Hz to 15000 Hz
50 Hz to 15000 Hz
300 Hz to 15000 Hz
6 Hz to 21000 Hz
50 Hz to 21000 Hz

BP17

-


Description of command

FW vers.

This command selects the first band pass in the AF analyzer to be used if the AF path coupling is set to DC (see CONFigure:AFANalyzer...:COUPling command).

V 3.05


Note:
In firmware versions ³V2.12 but <V3.05, the CONFigure:AFANalyzer... :FILTer:BPASs commands replace the CONFigure:AFANalyzer... :FILTer:BPASs:ACCoupling and CONFigure:AFANalyzer...:FILTer: BPASs:DCCoupling commands. No distinction is made between AC and DC path coupling. In firmware V3.00, no band pass selection is possible if DC path coupling is set. The CMU always uses a default band pass with a bandwidth of 6 Hz to 21000 Hz (BP17).


Home Previous Next