Reset Values of the Status Reporting Systems

The table below comprises the different commands and events causing the status reporting system to be reset. None of the commands, except for *RST and SYSTem:PRESet influences the functional instrument settings. In particular, DCL does not change the instrument settings.

Resetting instrument functions

Event Switching on
supply voltage
DCL,SDC      
  Power-On-Status-Clear (Device Clear, Selected Device Clear) *RST or SYSTem:PRESet STATus:PRESet *CLS
Effect 0 1        
Clear STB,ESR — yes — — — yes
Clear SRE,ESE — yes — — — —
Clear PPE — yes — — — —
Clear EVENt parts of the registers — yes — — — yes
Clear ENABle parts of all OPERation-and QUESTionable registers,
Fill ENABle parts of all other registers with "1".
— yes — — yes —
Fill PTRansition parts with “1"
Clear NTRansition parts
— yes — — yes —
Clear error queue yes yes — — — yes
Clear output buffer yes yes yes 1) 1) 1)
Clear command processing and input buffer yes yes yes — — —
 

1)Every command being the first in a command line, i.e. immediately following a <PROGRAM MESSAGE TERMINATOR> clears the output buffer.