Note:
Before returning to manual control, command processing must be completed.
If this is not the case, the CMU switches back to remote control immediately.
Returning to manual control by pressing any front panel key can be disabled
by the GPIB Local Lockout Message (LLO; see Chapter 8, Table Universal
Commands), which is also included in the NI commands SetRWLS (Set Remote
With Lockout State) or SendLLO. This prevents unintentional switchover,
i.e. return to manual control is possible via the GPIB bus only. The
CMU-specific command *LLO TRUE
is equivalent to LLO.
Returning to manual control via the front panel keys can be enabled again
by deactivating the REN control line of the GPIB bus (see GPIB
Bus Interface ). The CMU-specific command *LLO FALSe
also
enables return to manual control.