Reset/Cancel

Reset

The Reset button resets all settings in the selected function groups and test modes.

A box pops up to confirm the reset. While the reset is performed, the message Reset in progress is displayed. All running measurements are aborted and a connection to a DUT is dropped. Then the Reset popup window is closed and the CMU returns to the function group and test mode that was active when the reset was initiated.

Note:
A reset of the active function group is faster because no additional software modules must be loaded. Additive function groups (e.g. an Audio function group which complements a GSM function group) are reset together with the selected function groups.

A reset of all instrument settings can also be performed during the startup procedure.

The reset button restores the default values for manual control. In cases where the remote control commands use distinct default values (e.g. the repetition modes) only the manual default values are restored:

 

Remote control
SYSTem:PRESet[:ALL]
(base system)

 

Compare:
SYSTem:RESet[:ALL]
(default parameters in remote control, base system)
SYSTem:RESet:CURRent
(default param. in remote control, all function groups).

 

 

Cancel

The Cancel button cancels the selection that has been made and closes the menu. Cancel is selected by default when the Reset menu is opened.
 

Remote control
-