Home Previous Next


Display Mode

The Display Mode field defines which of the four measured and calculated measurement curves is displayed. The measurement curves differ in the way the burst power p(t) at a fixed point in time t is calculated if the measurement extends over several bursts:

Current Measured value for current burst
Minimum Minimum over a number of bursts
Maximum Maximum over a number of bursts
Average Average value over a number of bursts

The number of bursts for calculation of the statistical values Minimum, Maximum and Average – and thus the result – depends on the repetition mode set. In detail, this implies:

Single Shot Display of minimum, maximum and average value from the performed statistics cycle.
Continuous Display of minimum and maximum from all bursts already measured. The average value, however, is calculated according to the rule in Chapter 3, section General Settings.

Remote control
no display mode set explicitly, the four measurement curves are accessible via

FETCh:ARRAy:POWer[:CURRent]? FETCh:ARRAy:POWer:MINimum?
FETCh:ARRAy:POWer:MAXimum? FETCh:ARRAy:POWer:AVERage?
etc.


Home Previous Next