Level

The Level section defines the trigger thresholds if the measurement is triggered by the RF Power or IF Power (see Source) respectively. Both thresholds are defined relative to the maximum input level set in the Analyzer tab (see RF Max. Level softkey). The Level settings have no influence on Free Run or External trigger measurements.

Note:
The trigger levels are always relative to the current maximum input level. If RF Max. Level is set manually (RF Mode = Manual), the current input level is constant and equal to the setting value. In autoranging mode (RF Mode = Auto), the current maximum input level is dynamically adapted to the measured RF input level; the trigger levels change accordingly.

The RF Power trigger threshold is the RF input signal level (Wideband Power) beyond which the trigger condition is satisfied and a measurement is initiated.

Low

Medium

High

Low trigger threshold, equal to approx. the RF Max. Level –26 dB.

Medium trigger threshold, equal to approx. the RF Max. Level –16 dB.

High trigger threshold, equal to approx. the RF Max. Level –6 dB.

 

The IF Power trigger threshold is the IF trigger signal level beyond which the trigger condition is satisfied and a measurement is initiated. The IF Power input value defines the trigger threshold relative to the maximum input level:

IF power trigger threshold = <RF Max. Level> + <IF Power>

Remote control
TRIGger[:SEQuence]:THReshold:RFPower LOW | MEDium | HIGH

TRIGger[:SEQuence]:THReshold:IFPower <Power>