OK to Recall recalls the selected configuration file and activates the stored settings for the function groups and test modes selected in the configuration tree.
The active function group, test mode and measurement menu is stored with the Base system settings when a configuration file is created. As a consequence the behavior of the CMU after a recall depends on whether or not the Base system settings are also recalled:
Note:
Recalling only a configuration of the active function group is faster because
no additional software modules must be loaded.
Before recalling and activating a configuration file, the instrument checks
whether the settings are compatible with its current hardware configuration
and software versions. If an incompatibility is detected, the configuration
file is not recalled and an error message is generated. Configuration
files are upward compatible and can be re-used in later firmware versions.
Remote control
MMEMory:RECall:CURRent <FileName> [,<msus>]
MMEMory:RECall[:ALL] <FileName> [,<msus>]