Home Previous Next


Subarray Results

READ:SUBarrays:SPECtrum[:CURRent]?

READ:SUBarrays:SPECtrum:AVERage?

READ:SUBarrays:SPECtrum:MAXimum?

READ:SUBarrays:SPECtrum:MINimum?

Start single shot meas. and return results RUN

FETCh:SUBarrays:SPECtrum[:CURRent]?

FETCh:SUBarrays:SPECtrum:AVERage?

FETCh:SUBarrays:SPECtrum:MAXimum?

FETCh:SUBarrays:SPECtrum:MINimum?

Read meas. results (unsynchronized) RUN

SAMPle:SUBarrays:SPECtrum[:CURRent]?

SAMPle:SUBarrays:SPECtrum:AVERage?

SAMPle:SUBarrays:SPECtrum:MAXimum?

SAMPle:SUBarrays:SPECtrum:MINimum?

Read results (synchronized) RUN

Ret. values by subrange

Description of parameters

Def. value

Def. unit

-128.0 dBm to + 48.0 dBm
...
-128.0 dBm to + 48.0 dBm

Power[1], 1st value for power
...
Power[x], xth value for power

NAN
...
NAN

dBm
...
dBm


Description of command

FW vers.

These commands are always queries. They return the power versus frequency in the subranges defined by means of the CONFigure:SUBarrays:SPECtrum command. In the default setting of the configuration command the READ:SUBarrays..., FETCh:SUBarrays..., and SAMPle:SUBarrays... command group is equivalent to the READ:ARRay..., FETCh:ARRay..., and SAMPle:ARRay... command group described above.

The CONFigure:SUBarrays:SPECtrum command defines a maximum of 32 subranges. If one of the statistical modes (ARIThmetical, MINimum, MAXimum) or IVAL is set, only one value is returned by subrange.

The calculation of CURRent, AVERage, MINimum, and MAXimum results is explained in Chapter 3 (see Display Mode).

V 3.0


Home Previous Next