Home Previous Next


Measurement cycles

CONFigure:SPECtrum:CONTrol:REPetition <Repetition> ,<StopCondition> ,<Stepmode>

<Repetition>

Parameter description

Def. value

Def. unit

CONTinuous |
SINGleshot |

1 to 10000

Continuous measurement (until STOP or ABORT)
Single shot measurement (until Status = RDY)
Multiple measurement
(counting, until Status = STEP | RDY)

SING

-

<StopCondition>

Parameter description

Def. value

Def. unit

NONE

Continue measurement in case of error

NONE

-

<Stepmode>

Description of parameters

Def. value

Def. unit

STEP |
NONE

Interrupt measurement after each statistics cycle
Continue measurement according to its rep. mode

NONE

-


Description of command

FW vers.

This command defines the repetition mode and the stepping mode for the measurement. A stop condition is not available.

³V 1.15


Note:
In the case of READ commands (READ: to ) the <Repetition> parameter has no effect; the measurement is always stopped after a single shot.


Home Previous Next