Subarray Results

READ:SUBarrays:POWer[:CURRent]?
READ:SUBarrays:POWer:AVERage?
READ:SUBarrays:POWer:MAXimum?
READ:SUBarrays:POWer:MINimum?

Start single shot measurement and return results RUN

FETCh:SUBarrays:POWer[:CURRent]?
FETCh:SUBarrays:POWer:AVERage?
FETCh:SUBarrays:POWer:MAXimum?
FETCh:SUBarrays:POWer:MINimum?

Read meas. results (unsynchronized) RUN

SAMPle:SUBarrays:POWer[:CURRent]?
SAMPle:SUBarrays:POWer:AVERage?
SAMPle:SUBarrays:POWer:MAXimum?
SAMPle:SUBarrays:POWer:MINimum?

Read results (synchronized) RUN

Ret. values by subrange

Description of parameters

Def. value

Def. unit

-128.0 dBm to + 48.0 dBm
...
-128.0 dBm to + 48.0 dBm

Power[1], 1st value for power
...
Power[x], xth value for power

NAN
...
NAN

dBm
...
dBm

 

Description of command

FW vers.

These commands are always queries. They return the power versus time in the subranges defined by means of the CONFigure:SUBarrays:POWer command. In the default setting of the configuration command the READ:SUBarrays..., FETCh:SUBarrays..., and SAMPle:SUBarrays... command group is equivalent to the READ:ARRay..., FETCh:ARRay..., and SAMPle:ARRay... command group described above.

The CONFigure:SUBarrays:POWer command defines a maximum of 32 subranges. If one of the statistical modes (ARIThmetical, MINimum, MAXimum) or IVAL is set, only one value is returned by subrange.

V 3.0