Home Previous Next


Test cycles

CONFigure:NPOWer:CONTrol:REPetition <Repetition>,<StopCond>,<Stepmode>

<Repetition>

Parameter description

Def. value

Def. unit

CONTinuous |
SINGleshot |

1 ... 10000

Continuous measurement (until STOP or ABORT)
Single shot measurement (until Status = RDY)
Multiple measurement
(counting, until Status = STEP | RDY)

SING

-

<StopCondition>

Parameter description

Def. value

Def. unit

SONerror |
NONE

Start measurement in case of error (stop on error)
Continue measurement even in case of error

NONE

-

<Stepmode>

Description of parameters

Def. value

Def. unit

STEP |
NONE

Interrupt measurement after each statistics cycle
Continue measurement according to its rep. mode

NONE

-


Description of command

FW vers.

This command determines the repetition mode, the stop condition and the stepping mode for the measurement.

Note:

In the case of READ commands (READ:...), the <Repetition> parameter has no effect; the measurement is always stopped after a single shot.

³V 3.07


Home Previous Next