Home Previous Next


Disable Non Volatile RAM

SYSTem:NONVolatile:DISable

Description of command

FW vers.

This command has no query form. It prevents the CMU from saving measurement settings to the non volatile ram. This improves the system performance but implies that the current settings will not be saved for later sessions. Disabling the non volatile ram is recommended for instruments equipped with an FMR 6 Front Module controller that are operated in remote control mode.

³V 3.0

 

Note:
There is no way to cancel the effect of the SYSTem:NONVolatile:DISable command within the current session, even is the CMU is switched to manual control (Local). To re-enable the non volatile ram, the CMU must be rebooted.


Home Previous Next