Ref. Frequency Source

CONFigure:SYNChronize:FREQuency:REFerence:MODE <Mode>

<Mode>

Parameter description

Def. value

Default unit

INTernal |
EXTernal

Internal reference frequency used
External reference frequency used

INT

-


Description of command

FW vers.

The command defines the source of the synchronization signal. After activating the external reference frequency (e.g. after a reset of the base system where the reference frequency is set to INTernal) it is necessary to allow for a setting time (~1 s) until the CMU has synchronized. The query [SENSe:]SYNChronize:FREQuency:REFerence:LOCKed? indicates whether the reference frequency is locked. A partial reset of all function groups with the exception of the base system does not reset the source of the reference frequency.

³V 1.12