Home Previous Next


Subarray Results

READ:SUBarrays:MULTitone:AF1Channel?

READ:SUBarrays:MULTitone:AF2Channel?

Start single shot measurement and return results RUN

FETCh:SUBarrays:MULTitone:AF1Channel?

FETCh:SUBarrays:MULTitone:AF2Channel?

Read meas. results (unsynchronized) RUN

SAMPle:SUBarrays:MULTitone:AF1Channel?

SAMPle:SUBarrays:MULTitone:AF2Channel?

Read results (synchronized) RUN

Ret. values per subrange

Description of parameters

Def. value

Def. unit

-100.0 dB to +20.0 dB,
...
-100.0 dB to +20.0 dB

FreqResp[1], 1st value for AF response
...
FreqResp[n], nth value for AF response

NAN
...
NAN

dB
...
dB

 

Description of command

FW vers.

These commands are always queries. They return the audio level in the subranges defined by means of the CONFigure:SUBarrays:MULTitone:AF1Channel command. In the default setting of the configuration command the READ:SUBarrays..., FETCh:SUBarrays..., and SAMPle:SUBarrays... command group is equivalent to the READ:ARRay..., FETCh:ARRay..., and SAMPle:ARRay... command group described above.

The CONFigure:SUBarrays:MULTitone:AF1Channel command defines a maximum of 32 subranges. If one of the statistical modes (ARIThmetical, MINimum, MAXimum) or IVAL is set, only one value is returned by subrange.

V 3.0


Home Previous Next