Home Previous Next


Scalar Results

READ[:SCALar]:AFANalyzer[:PRIMary]?

READ[:SCALar]:AFANalyzer:SECondary?

Start single shot meas. and return results

FETCh[:SCALar]:AFANalyzer[:PRIMary]?

FETCh[:SCALar]:AFANalyzer:SECondary?

Read out meas. results (unsynchronized)

SAMPle[:SCALar]:AFANalyzer[:PRIMary]?

SAMPle[:SCALar]:AFANalyzer:SECondary?

Read out meas. results (synchron.)

Return

Description of parameters

Def. value

Def. unit

PeakVoltage1
RMSVoltage1
DCVoltage
Distortion

PeakVoltage2
RMSVoltage2

Frequency

0 V to 42.4 V
0 V to 30 V
-30 V to 30 V
0% to 100%
0 V to 42.4 V
0 V to 30 V
10 Hz to 204.8 kHz

NAN
NAN
NAN
NAN
NAN
NAN
NAN

V
V
V
%
V
V
Hz

 

Description of command

FW vers.

These commands are always queries. They start a measurement and output all scalar measurement results (see also Subsystem AFANalyzer...:FILTer (Filter)). These are:

  • Peak1 and RMS1 value of AC voltage after first band pass and the weighting filter
  • DC voltage
  • Total harmonic distortion
  • Peak2 and RMS2 value of AC voltage after first band pass and second band pass
  • (variable band pass)
  • Frequency counter