Definition of Subarrays

CONFigure:SUBarrays:MULTitone:AF1Channel

CONFigure:SUBarrays:MULTitone:AF2Channel

<Mode>,<Start>,<Samples>{,<Start>,<Samples>}

<Mode>

Description of parameters

Def. value

Def. unit

ALL |
ARIThmetical |
MINimum |
MAXimum |
IVAL,

Return all measurement values
Return arithm. mean value in every range
Return minimum value in every range
Return maximum value in every range
Return single interpolated value at <Start>

ALL

-

<Start>

Description of parameters

Def. value

Def. unit

1 to 20,

Start test tone in current range

1

-

<Samples>

Description of parameters

Def. value

Def. unit

1 to 20

Number of test tones in current range

20

-


Description of command

FW vers.

This command configures the READ:SUBarrays:MULTitone:AF1Channel..., FETCh:SUBarrays:MULTitone:AF1Channel..., and SAMPle:SUBarrays:MULTitone:AF1Channel commands. It restricts the measurement to up to 32 subranges where either all measurement results (the number of which is given by the <Samples> parameter) or a single statistical value is returned.

For <Mode> = IVAL, the <Samples> parameter is ignored and the CMU returns a single measurement value corresponding to the abscissa value <Start>. If <Start> is located between two test points with valid results then the result is calculated from the results at these two adjacent test points by linear interpolation.

The subranges are subsets of the full range of test tones defined via CONFigure:MULTitone:AF1Channel:TONE<nr>. Each subrange contains all test tones between the start test tone (test tone no. <Start>) and test tone no.
<Start>+<Samples>-1. Test points inside this range that are disabled are not measured (result NAN) and do not enter into the ARIThmetical, MINimum and MAXimum values.

By default, only one range corresponding to the total measurement range is used and all measurement values are returned.

V 3.0