CONFigure:AFANalyzer[:PRIMary]:CONTrol:REPetition
CONFigure:AFANalyzer:SECondary:CONTrol:REPetition
<Repetition>, <StopCondition>, <Stepmode>
<Repetition> |
Description of parameters |
Def. value |
Def. unit |
CONTinuous | |
Continuous measurement (until |
SING |
- |
<StopCondition> |
Description of parameters |
Def. value |
Def. unit |
NONE |
Continue measurement even in case of error |
NONE |
- |
<Stepmode> |
Description of parameters |
Def. value |
Def. unit |
STEP | |
Interrupt measurement after each statistic cycle Continue measurement according to its rep. mode |
NONE |
- |
Description of command |
FW vers. |
This command determines the number of statistics cycles and the stepping mode for the measurement. A stop condition is not available. |
³V 2.10 |
Note:
In the case of READ commands (READ
: to) the <Repetition> parameter has no effect; the measurement is always stopped after a single shot.