The CMU can be connected to an external controller via the GPIB bus (IEEE bus according to standard IEEE 488; throughout this documentation we will primarily use the term GPIB bus which is also used in the operating menus and in the SCPI command syntax) or via serial interface:
Connection via GPIB bus
|
The CMU is connected to the GPIB interface of the controller via the GPIB bus connector (IEEE 488 / IEC 625) at the rear of the instrument and a shielded cable. The technical specifications of the GPIB interface are listed in section "Hardware Interfaces " in Chapter 8. |
GPIB Bus Configuration |
In the default configuration the CMU accepts commands from either the GPIB or COM 1 interface. The parameters for GPIB bus control of the CMU are set in the Remote tab of the Setup popup menu (in the following abbreviated by Setup – Remote, see also chapter 4, Settings for Remote Control). Ø To open the Setup - Remote menu, press the SETUP key at the front of the instrument and activate the Remote hotkey at the lower edge of the screen. Ø Use the rotary knob to move the focus onto the SCPI Connection section of the Setup table. If necessary, press the rotary knob or the ON/OFF key to expand the parameters in the table (see Chapter 3). Ø In the Port table row select either GPIB + Com 1 or GPIB bus interface for transmission. The bus address is factory-set to 20. It can be changed in the Primary Address input field.
|
Connection via serial interface
|
The CMU can be connected to the serial interface of a controller via one of the serial interfaces COM 1 or COM 2 and a so-called null-modem cable. The pin assignment and wiring of a null-modem cable are described in section Handshake of chapter 8. The technical specifications of the serial (RS-232-C) interface are also discussed in chapter 8 (refer to section Hardware Interfaces). Either a 25-pin or a 9-pin connector can be used on the controller side. It may be necessary to use an appropriate adapter (see chapter 8, Hardware Interfaces). |
Selection
|
In the default configuration the CMU accepts commands from either the GPIB or COM 1 interface. The COM 2 interface must be selected explicitly. Ø Proceed as described above to activate the Remote tab of the Setup menu. Ø In the Port table row, select GPIB + Com 1 or COM 1 or COM 2 to activate one of the RS-232 interfaces for data transfer.
|
Configuration
|
After selection of a serial interface, the transmission parameters must be set to comply with the parameters of the addressed device. This is done in the Comm. (communications) tab of the Setup menu:
Ø To open the Setup – Comm. tab press the SETUP key at the front of the instrument and activate the Comm. hotkey at the lower edge of the screen. Ø In the table section corresponding to the selected COM port check the settings for the Baudrate, Data Bits, Parity, and Protocol.
|